Ocean Optics offers advanced Raman probes designed to meet the demands of both research and industrial environments. Now their range has been extended with the release of their Dual Wavelength Raman Probe featuring a colinear optical design and high-throughput optics for efficient, simultaneous measurements at 532/785 nm—ideal for complex sample interrogation. Also available is the new Industrial Process Raman Probe is built for rugged applications, with a sealed, swageable lens shaft suitable for immersion, pressure, and vacuum conditions.

Both probes incorporate backscattering optics for high collection efficiency and are compatible with a wide range of sample types including solids, liquids, and gases.

Specifications ORP-DW-532-785 ORP-P785-6 ORP-XXX
Excitation Wavelength 532/785 nm 785 nm 532 nm, 638 nm, 785 nm, 830 nm, 1064 nm
Spectral Range 100-4000 cm-1 (The ultimate range is spectrograph/detector dependent) 100-4000 cm-1 (The ultimate range is spectrograph/detector dependent) 100-4000 cm-1 (The ultimate range is spectrograph/detector dependent)
Spot Size at the Sample ~100 microns for 100 µ core excitation optical fiber 100 µ for standard optical fiber (fiber core dependent) 100 µ for standard optical fiber (fiber core dependent)
Focal Length 9 mm standard (12,15, & 18 mm optional)
Note: Probe efficiency decreases with increasing focal length.
9 mm standard (12,15, & 18 mm optional)
Note: Probe efficiency decreases with increasing focal length.
9 mm standard (12,15, & 18 mm optional)
Note: Probe efficiency decreases with increasing focal length.
Working Distance NA 7 mm for standard lens 7 mm for standard lens
Numerical Aperture NA 0.22 with standard lens 0.22 with standard lens
Focusing Lens Barrel Non-immersion or sealed probe shaft NA NA
Probe Body Dimensions 3.8” length x 1.5” diameter 4.5” length x 1.3” diameter 2.25” L x 0.96” W x 0.58” H
Probe Body Material Hard anodized aluminum Hard anodized aluminum Hard anodized aluminum
Probe Body Seal NA Buna-N O-ring NA
Probe Shaft Dimensions 3/8”diameter x 6” length 3/8” diameter x 2” length 3/8” diameter x 2” length (custom lengths available)
Probe Shaft Tip Seal NA Kalrez® O-ring NA
Probe Shaft Material 316 stainless steel
(other metals such as Hastelloy and Inconel are available)
316 stainless steel
(other metals are available)
316 stainless steel
Probe Shaft Window NA Sapphire (Fused silica available) NA
Filter Efficiency OD > 6 at laser wavelength OD > 6 at laser wavelength OD > 6 at laser wavelength
Operating Temperature 0 – 85 ⁰C (non-immersion shaft)
0 – 325 ⁰C (sealed shaft)
0 – 325 ⁰C 0 – 85 ⁰C
Maximum Operating Pressure 15 psi (non-immersion shaft)
6000 psi (sealed shaft)
6000 psi 15 psi
Fiber Configuration 100/200 µ core standard
Note: Custom optical fiber cores are available.
105/600 µ core standard to laser and spectrometer respectively
Note: Custom optical fiber cores are available.
100/600 µ core standard to laser and spectrometer respectively
Note: Custom optical fiber cores are available.
Fiber Optic Cable 3 m reinforced stainless steel armor cable standard
Note: Custom lengths are available.
3 m reinforced stainless steel armor cable standard
Note: Custom lengths are available.
1 m reinforced stainless steel armor cable standard
Note: Custom lengths are available.
Coupling System Laser (FC), Spectrometer (SMA) Laser (FC), Spectrometer (SMA) Laser (FC), Spectrometer (SMA)